System Suitability Testing: Simplifying A Critical Element In The TOC Monitoring Process
White Paper: System Suitability Testing Simplifying A Critical Element In The TOC Monitoring Process
In the Pharmaceutical Water production process, System Suitability Testing (SST) is an essential activity to verify the performance of a Total Organic Carbon monitoring system and to ensure its adequacy for TOC analysis.
USP and EP Requirements
In the requirements for TOC measurement,
The United States Pharmacopoeia
and European Pharmacopoeia have established
specific Total Organic Carbon
(TOC) tests as described in USP General
Chapter ‹643› and EP Chapter 2.2.44, respectively.
These chapters provide:
- Guidance on the methodology for TOC testing,
- Criteria for establishing instrument acceptance, and
- TOC limits for the sample to be tested.
The types of water subject to TOC testing are:
- USP Purified Water
- USP Water for Injection
- EP Aqua Purificata (Purified Water) – TOC may be performed in lieu of oxidizable substances test
- EP Valde Aqua Purificata (Highly Purified Water)
- EP Aqua ad Injectabilia (Water for Injections).
- JP Water for Injection, effective April, 2006
Challenges of System
Suitability Testing
Because organic carbon appears in various
forms in nature and subsequently in
water treatment processes, a wide variety
of oxidation states and chemical forms are
found in these systems. The goal of System
Suitability Testing is to challenge the TOC
measurement technique by verifying that
two chemicals of very different chemical
properties respond equally. In this case,
the two chemicals specified in the pharmacopoeial
chapters are sucrose and 1,4-
benzoquinone. Because of their unique
and different chemical structure, sucrose
and 1,4-benzoquinone challenge the
bond-breaking and oxidation capability of
the TOC measurement technology. These
solutions are commonly referred to as easy
to oxidize and hard to oxidize chemicals,
respectively. In addition to sucrose and 1,4
benzoquinone, reagent water (the water
used to manufacture the sucrose and 1,4
benzoquinone solutions) is analyzed during
the suitability testing.